Specifications
Calibrated DyeSYBR™ Green I, ROX™, Texas Red™, NED™, JOE™, Cy3™, Cy5™, VIC™, FAM™, TAMRA™
DimensionsWidth 34 cm (13.39 in.), Depth 45 cm (17.72 in.), Height 49 cm (19.29 in.)
Dynamic Range9 Logs
Electrical Requirements950 W
External ComputerDell Notebook D520
For Use With (Equipment)7500 System
Format96-well Plate
High-throughput CompatibilityMultiplexing
OpticsCCD Camera, 5 Excitation Filters, Tungsten-halogen Lamp, 5 Emission Filters
Passive Reference DyeROX (Separate Tube), No ROX, ROX (Pre-mixed)
Peak Block Ramp Rate2.5°C⁄s
Precision99.7% Confidence (5000 & 10000 Copies)
Quantity1 system
Reaction SpeedStandard
Reaction Volume Range20-100 μl
Sample Ramp Rate± 1.1°C⁄s
Sensitivity1 Copy of RNase P gene (Human Genomic DNA)
Software CompatibilityGeneMapper™ ID v3.2, 7500 System Software, HID Real Time PCR Analysis Software v1.1
Temperature Range (Metric)4 to 100°C
Thermal Accuracy0.25°C (35°C to 95°C after 3 min.)
Thermal Cycling SystemPeltier-based System
Thermal Uniformity0.50°C (after 30 s)
Throughput4 to 5-96-well Plates per 8-hr day
Voltage120 V
Weight75 lb.
Unit SizeEach