Specifications
Product TypeSOLA iQ On-line Sulfur Analyzer
Ambient Temperature12°C to 40°C
Detection Limits1.25% of full scale or 125 ppb (whichever is greater)
Detector TypePulsed UV Fluorescence (PUVF) with Pyrolyzer for Total Sulfur Measurement
Voltage100/240 V
Certifications/ComplianceNEC Class I, Division 2, Groups B, C & D. T2/T3 & T4 (T4 rating is application dependent)
NEC Class I, Division 1, Groups B, C & D. T2/T3 & T4 (T4 rating is application dependent)
CSA with associated “C/US Mark” Class I, Division 2, Groups B, C & D. T2/T3 & T4 (T4 rating is application dependent)
CSA with associated “C/US Mark” Class I, Division 1, Groups B, C & D. T2/T3 & T4 (T4 rating is application dependent)
ATEX Zone 1 Ex px IIC T3
ATEX Zone 2 Ex pz IIC T3 IECEx Zone 1 Ex px IIC T3
Communication DetailsAutoCONFIG™ software: Security protected access to all analyzer setup and configuration parameters and all process and diagnostic data. Connect via a standard PC across your Local Area Network. Download 30,000 data records (process and analzer functions), typically up to 24 hours of records (user configurable).
Current OutputsRelay & digital outputs
8x relay outputs rated 6A at 240VAC
8x solid state relay outputs rated 0.2A at 120VAC/VDC
Depth (English)18 in.
Depth (Metric)459 mm
DisplayFront panel mounted 7 in. color touchscreen user interface to access analyser functions and diagnostic data including sulfur concentrations, oven and furnace temperatures, PMT and lamp voltages, detector flow rate and more.
Electrical Requirements100/240 V, 50/60 Hz, 18 A circuit recommended, 18 A maximum during warm-up cycle, 7 to 8 A once achieving operational temperatures
Height (English)Zone 1: 56 in., Zone 2: 45 in.
Height (Metric)Zone 1: 1420 mm, Zone 2: 1130 mm
Hertz50/60 Hz
Inputs4x inputs (user configurable as 4 to 20 mA or 0-5V)
LinearityRanges ≥10 ppm: ±2% of full scale, one sample injection per minute; ±1% of full scale, two sample injections per minute
Ranges ≤10 ppm: ±2% of full scale, two sample injections per minute
Measurement RangesSOLA iQ Liquid: Full scale ranges from 5ppm to 5% - single or dual range analyzer
SOLA iQ Vapor: Full scale ranges from 5ppm to 1% - single or dual range analyzer
SOLA iQ Condensible Vapors (CV): Full scale ranges from 5ppm to 1% - single or dual range analyzer
SOLA iQ Flare: Full scale ranges from 10ppm to 100% - dual range analyzer
SOLA iQ Trace: Full scale ranges from 2ppm to 50ppm - single or dual range analyzer
No. of Process StreamsSingle or multiple (up to 4) stream control via SOLA iQ pneumatic outputs, stream selector hardware to be supplied by customer/third party
Outputs4x 4 to 20 mA outputs
PrecisionSOLA iQ Liquid & Vapor: Ranges ≥10ppm ±2% of full scale, 1 sample injection per minute; ±1% of full scale, 2 sample injections per min
SOLA iQ Liquid & Vapor: Ranges <10ppm ±2% of full scale, two sample injections per minute
SOLA iQ Flare: ±1% of full scale, two sample injections per minute
SOLA iQ Trace: ±1% of full scale, two sample injections per minute
Response TimeEvery 1 second, typically 5-6 minutes to 90% of new value (application dependent)
Serial ConnectionsRS232 Serial Port
Width (English)24 in.
Width (Metric)610 mm
Zero Grade Air3.8 to 6.9 barg, 180 to 210 L/min. (application dependent), Oil Free, -40°C (-40°F) dew point
Unit SizeEach