Get near-tissue equivalent thermoluminescence dosimetry (TLD) performance with a high-capacity reader for all types of TLD elements and health physics applications. The Thermo Scientific™ HARSHAW TLD Model 5500 Automatic Reader provides unattended operation for up to 50 dosimeters and accommodates TL chips, disks, rods and cubes in a broad variety of sizes. This reader is suitable for all kinds of applications that require TLD detectors. Clinical applications are: Radiotherapy planning verification, total body irradiation dose verification, critical organ dose verification, diagnostic dose studies including CT, and many others where measurement quality assurance is critical.
Fuente de alimentación | Selectable, 110/120 VAC ±10%, 60 Hz or 220/240 VAC ±10%, 50 Hz |
Perfiles de tiempo y temperatura | Multiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles |
Uniformidad de lote | ±30% from the factory standard |
Descripción | Model 5500 TLD Reader Instrument, Macro Pick Version |
Intervalo de dosis | Measurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear) Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter |
Capacidad de reutilización del dosímetro | >500 |
Intervalo dinámico | 7 decades |
Caudal | Flow Rate: internally regulated to 400 liters/h if specified pressure is provided |
Altura (imperial) | 1.89 in. |
Altura (métrico) | 48 cm |
Humedad | Functions within specifications after 24-hour exposure to 95% humidity and subsequent 6-hour recovery with the use of nitrogen supply |
Linealidad | Better than 1% |
PMT | Thermoelectric PMT cooler for Maximum gain stability |
Tipo de radiación | Photons, Electrons, Neutrons |
Luz de referencia | Temperature-stabilized LED |
Repetibilidad | for 1 mGy (100 mrad) 137 Cs doses, >2% standard deviation of 10 sequential measurements |
Estabilidad | HV stability: ±0.005% per hour, ±0.02% per 8 hours |
Temperatura de funcionamiento (imperial) | 59°F to 104°F |
Temperatura de almacenamiento (imperial) | 14°F to 140°F |
Temperatura de funcionamiento (métrico) | 15°C to 40°C |
Temperatura de almacenamiento (métrico) | -10°C to +60°C |
Peso (imperial) | 88.18 lb. |
Peso (métrico) | 40 kg |
Exposición a la luz ambiental | Functions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader |
Para utilizar con (equipo) | Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories |
Tipo de producto | Radiation Dosimeter Reader |
Tipo de detector | Photomultiplier Tube |
Sensibilidad | <10μGy |
Unit Size | Each |
Fuente de alimentación | 100/120/220/240 VAC |
Perfiles de tiempo y temperatura | Multiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles; Typical cycle time per detector: typically 30 seconds |
Uniformidad de lote | ±30% from the factory standard |
Descripción | Model 5500 TLD Reader Instrument |
Intervalo de dosis | Measurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear) Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter |
Capacidad de reutilización del dosímetro | >500 |
Intervalo dinámico | 7 decades |
Caudal | Flow Rate: internally regulated to 400 liters/h if specified pressure is provided |
Altura (imperial) | 1.89 in. |
Altura (métrico) | 48 cm |
Humedad | Functions within specifications after 24-hour exposure to 95% and subsequent 6-hour recovery |
Linealidad | Better than 1% |
PMT | Thermoelectric PMT cooler for Maximum gain stability |
Tipo de radiación | Photons, Electrons, Neutrons |
Luz de referencia | Temperature-stabilized LED |
Repetibilidad | for 1 mGy (100 mrad) 137 Cs doses, >2% standard deviation of 10 sequential measurements |
Estabilidad | HV stability: ±0.005% per hour, ±0.02% per 8 hours |
Temperatura de funcionamiento (imperial) | 59°F to 104°F |
Temperatura de almacenamiento (imperial) | 14°F to 140°F |
Temperatura de funcionamiento (métrico) | 15°C to 40°C |
Temperatura de almacenamiento (métrico) | -10°C to +60°C |
Peso (imperial) | 88.18 lb. |
Peso (métrico) | 40 kg |
Exposición a la luz ambiental | Functions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader |
Para utilizar con (equipo) | Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories |
Tipo de producto | Radiation Dosimeter Reader |
Tipo de detector | Thermoluminescence (TL) Detector |
Sensibilidad | Dependent on config. Used |
Unit Size | Each |
Descripción | 3 Year Extended Depot Service Warranty |
Tipo de radiación | Photons, Electrons, Neutrons |
Tipo de producto | Radiation Dosimeter Reader |
Tipo de detector | Thermoluminescence (TL) Detector |
Sensibilidad | <10μGy |
Unit Size | Each |
Número de catálogo | Especificaciones | Formato | Descripción | Precio (ARS) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
5500READERMA | Each | Model 5500 TLD Reader Instrument, Macro Pick Version | Contáctenos › | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
HARSHAWTLD5500 | Each | Model 5500 TLD Reader Instrument | Contáctenos › | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5500TLD-DS3 | Each | 3 Year Extended Depot Service Warranty | Contáctenos › | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
The HARSHAW TLD Model 5500 Manual Reader includes a carrier disk for up to 50 TL detectors, a linear, programmable contactless nitrogen heating system, and a cooled photomultiplier tube with associated electronics to measure the TL light output. The operational software, Thermo Scientific™ WinREMS™, which runs on a separate computer, provides the user interface, the reader control and the applications software.
Dosimetry Performance Using LiF:Mg,Ti (TLD-100/600/700):
Instrument Performance:
Ordering Alerts:
Computer requirements: Any commercially available PC system with Windows™ 7, XP, 2000, or NT 4. 32 bit OS only. RS232 strongly recommended. (Please contact your sales representative for approved USB to RS232 converters).
Compatible with:
Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories
Recommended for:
Treatment planning verification, diagnostic dose studies, quality assurance, whole body dosimetry, extremity dosimetry, environmental dosimetry, high dose applications (radiation hardness testing), and many more.