Thermo Scientific™

Harshaw TLD Model 5500 Readers

Número de catálogo: HARSHAWTLD5500
Thermo Scientific™

Harshaw TLD Model 5500 Readers

Número de catálogo: HARSHAWTLD5500

Get near-tissue equivalent thermoluminescence dosimetry (TLD) performance with a high-capacity reader for all types of TLD elements and health physics applications. The Thermo Scientific™ HARSHAW TLD Model 5500 Automatic Reader provides unattended operation for up to 50 dosimeters and accommodates TL chips, disks, rods and cubes in a broad variety of sizes. This reader is suitable for all kinds of applications that require TLD detectors. Clinical applications are: Radiotherapy planning verification, total body irradiation dose verification, critical organ dose verification, diagnostic dose studies including CT, and many others where measurement quality assurance is critical.

 
Número de catálogo
HARSHAWTLD5500
Formato
Each
Descripción
Model 5500 TLD Reader Instrument
Precio (ARS)
Especificaciones completas
Fuente de alimentación100/120/220/240 VAC
Perfiles de tiempo y temperaturaMultiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles; Typical cycle time per detector: typically 30 seconds
Uniformidad de lote±30% from the factory standard
DescripciónModel 5500 TLD Reader Instrument
Intervalo de dosisMeasurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear)
Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter
Capacidad de reutilización del dosímetro>500
Intervalo dinámico7 decades
CaudalFlow Rate: internally regulated to 400 liters/h if specified pressure is provided
Altura (imperial)1.89 in.
Altura (métrico)48 cm
HumedadFunctions within specifications after 24-hour exposure to 95% and subsequent 6-hour recovery
LinealidadBetter than 1%
PMTThermoelectric PMT cooler for Maximum gain stability
Tipo de radiaciónPhotons, Electrons, Neutrons
Luz de referenciaTemperature-stabilized LED
Repetibilidadfor 1 mGy (100 mrad) 137 Cs doses, >2% standard deviation of 10 sequential measurements
EstabilidadHV stability: ±0.005% per hour, ±0.02% per 8 hours
Temperatura de funcionamiento (imperial)59°F to 104°F
Temperatura de almacenamiento (imperial)14°F to 140°F
Temperatura de funcionamiento (métrico)15°C to 40°C
Temperatura de almacenamiento (métrico)-10°C to +60°C
Peso (imperial)88.18 lb.
Peso (métrico)40 kg
Exposición a la luz ambientalFunctions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader
Para utilizar con (equipo)Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories
Tipo de productoRadiation Dosimeter Reader
Tipo de detectorThermoluminescence (TL) Detector
SensibilidadDependent on config. Used
Unit SizeEach
Número de catálogo
Formato
Each
Descripción
3 Year Extended Depot Service Warranty
Precio (ARS)
Especificaciones completas
Descripción3 Year Extended Depot Service Warranty
Tipo de radiaciónPhotons, Electrons, Neutrons
Tipo de productoRadiation Dosimeter Reader
Tipo de detectorThermoluminescence (TL) Detector
Sensibilidad<10μGy
Unit SizeEach
Número de catálogo
Formato
Each
Descripción
Model 5500 TLD Reader Instrument, Macro Pick Version
Precio (ARS)
Especificaciones completas
Fuente de alimentaciónSelectable, 110/120 VAC ±10%, 60 Hz or 220/240 VAC ±10%, 50 Hz
Perfiles de tiempo y temperaturaMultiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles
Uniformidad de lote±30% from the factory standard
DescripciónModel 5500 TLD Reader Instrument, Macro Pick Version
Intervalo de dosisMeasurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear)
Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter
Capacidad de reutilización del dosímetro>500
Intervalo dinámico7 decades
CaudalFlow Rate: internally regulated to 400 liters/h if specified pressure is provided
Altura (imperial)1.89 in.
Altura (métrico)48 cm
HumedadFunctions within specifications after 24-hour exposure to 95% humidity and subsequent 6-hour recovery with the use of nitrogen supply
LinealidadBetter than 1%
PMTThermoelectric PMT cooler for Maximum gain stability
Tipo de radiaciónPhotons, Electrons, Neutrons
Luz de referenciaTemperature-stabilized LED
Repetibilidadfor 1 mGy (100 mrad) 137 Cs doses, >2% standard deviation of 10 sequential measurements
EstabilidadHV stability: ±0.005% per hour, ±0.02% per 8 hours
Temperatura de funcionamiento (imperial)59°F to 104°F
Temperatura de almacenamiento (imperial)14°F to 140°F
Temperatura de funcionamiento (métrico)15°C to 40°C
Temperatura de almacenamiento (métrico)-10°C to +60°C
Peso (imperial)88.18 lb.
Peso (métrico)40 kg
Exposición a la luz ambientalFunctions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader
Para utilizar con (equipo)Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories
Tipo de productoRadiation Dosimeter Reader
Tipo de detectorPhotomultiplier Tube
Sensibilidad<10μGy
Unit SizeEach
Mostrando 3 de 3
Número de catálogoEspecificacionesFormatoDescripciónPrecio (ARS)
HARSHAWTLD5500Especificaciones completas
EachModel 5500 TLD Reader InstrumentContáctenos ›
Fuente de alimentación100/120/220/240 VAC
Perfiles de tiempo y temperaturaMultiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles; Typical cycle time per detector: typically 30 seconds
Uniformidad de lote±30% from the factory standard
DescripciónModel 5500 TLD Reader Instrument
Intervalo de dosisMeasurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear)
Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter
Capacidad de reutilización del dosímetro>500
Intervalo dinámico7 decades
CaudalFlow Rate: internally regulated to 400 liters/h if specified pressure is provided
Altura (imperial)1.89 in.
Altura (métrico)48 cm
HumedadFunctions within specifications after 24-hour exposure to 95% and subsequent 6-hour recovery
LinealidadBetter than 1%
PMTThermoelectric PMT cooler for Maximum gain stability
Tipo de radiaciónPhotons, Electrons, Neutrons
Luz de referenciaTemperature-stabilized LED
Repetibilidadfor 1 mGy (100 mrad) 137 Cs doses, >2% standard deviation of 10 sequential measurements
EstabilidadHV stability: ±0.005% per hour, ±0.02% per 8 hours
Temperatura de funcionamiento (imperial)59°F to 104°F
Temperatura de almacenamiento (imperial)14°F to 140°F
Temperatura de funcionamiento (métrico)15°C to 40°C
Temperatura de almacenamiento (métrico)-10°C to +60°C
Peso (imperial)88.18 lb.
Peso (métrico)40 kg
Exposición a la luz ambientalFunctions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader
Para utilizar con (equipo)Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories
Tipo de productoRadiation Dosimeter Reader
Tipo de detectorThermoluminescence (TL) Detector
SensibilidadDependent on config. Used
Unit SizeEach
5500TLD-DS3Especificaciones completas
Each3 Year Extended Depot Service WarrantyContáctenos ›
Descripción3 Year Extended Depot Service Warranty
Tipo de radiaciónPhotons, Electrons, Neutrons
Tipo de productoRadiation Dosimeter Reader
Tipo de detectorThermoluminescence (TL) Detector
Sensibilidad<10μGy
Unit SizeEach
5500READERMAEspecificaciones completas
EachModel 5500 TLD Reader Instrument, Macro Pick VersionContáctenos ›
Fuente de alimentaciónSelectable, 110/120 VAC ±10%, 60 Hz or 220/240 VAC ±10%, 50 Hz
Perfiles de tiempo y temperaturaMultiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles
Uniformidad de lote±30% from the factory standard
DescripciónModel 5500 TLD Reader Instrument, Macro Pick Version
Intervalo de dosisMeasurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear)
Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter
Capacidad de reutilización del dosímetro>500
Intervalo dinámico7 decades
CaudalFlow Rate: internally regulated to 400 liters/h if specified pressure is provided
Altura (imperial)1.89 in.
Altura (métrico)48 cm
HumedadFunctions within specifications after 24-hour exposure to 95% humidity and subsequent 6-hour recovery with the use of nitrogen supply
LinealidadBetter than 1%
PMTThermoelectric PMT cooler for Maximum gain stability
Tipo de radiaciónPhotons, Electrons, Neutrons
Luz de referenciaTemperature-stabilized LED
Repetibilidadfor 1 mGy (100 mrad) 137 Cs doses, >2% standard deviation of 10 sequential measurements
EstabilidadHV stability: ±0.005% per hour, ±0.02% per 8 hours
Temperatura de funcionamiento (imperial)59°F to 104°F
Temperatura de almacenamiento (imperial)14°F to 140°F
Temperatura de funcionamiento (métrico)15°C to 40°C
Temperatura de almacenamiento (métrico)-10°C to +60°C
Peso (imperial)88.18 lb.
Peso (métrico)40 kg
Exposición a la luz ambientalFunctions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader
Para utilizar con (equipo)Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories
Tipo de productoRadiation Dosimeter Reader
Tipo de detectorPhotomultiplier Tube
Sensibilidad<10μGy
Unit SizeEach
Mostrando 3 de 3

The HARSHAW TLD Model 5500 Manual Reader includes a carrier disk for up to 50 TL detectors, a linear, programmable contactless nitrogen heating system, and a cooled photomultiplier tube with associated electronics to measure the TL light output. The operational software, Thermo Scientific™ WinREMS™, which runs on a separate computer, provides the user interface, the reader control and the applications software.

  • Multiple, programmable, linear heating cycles including pre-read anneal, acquire and post-read anneal cycles.
  • Typical cycle time per detector: typically 30 seconds.
  • Heating temperature capability up to 400°C.
  • Thermoelectric PMT cooler for maximum gain stability.
  • Seven decade glow curve acquisition range.
  • Options include glow curve deconvolution software (WinGCF Software), various types of neutral density filter to extend the high measurement range, and TL element specific carrier disks for best possible reproducibility.
  • Warm-up time: 30 minutes.
  • Easy to operate, service and maintain.
  • Size: 38 × 46 × 48cm.
  • 40kg.

Dosimetry Performance Using LiF:Mg,Ti (TLD-100/600/700):

  • Nearly tissue equivalent.
  • Radiation types: photons (E>5keV), electrons (E>70keV), neutrons (from thermal energies up to ∽100MeV, TLD-600 only).
  • Measurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear).
  • Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter.
  • Repeatability: Less than 2% variation based on one standard deviation for 10 sequential measurements of 1 mGy 137-Cs.
  • Residual TL signal: <0.2% of reading.
  • Fading: <20% in 3 months without thermal treatment; <5% in 3 months using pre-read anneal or glow curve deconvolution software (WinGCF Software).
  • Batch uniformity: ±30% from the factory standard.
  • Reuse: typically %500.

Instrument Performance:

  • Dynamic range: 7 decades.
  • HV stability: ±0.005% per hour, ±0.02% per 8 hours.
  • Linearity: better than 1%.
  • Test light: temperature-stabilized LED.
  • Electrical supply: selectable, 110/120VAC ±10%, 60Hz or 220/240VAC ±10%, 50Hz
  • Nitrogen supply:
    Purified (99.995%), dry.
    Pressure: 2.8 to 6.3kg/cm2.
    Flow Rate: internally regulated to 400 liters/h if specified pressure is provided.
  • Operating temperature: 15° to 40°C.
  • Storage temperature: -10° to +60°C.
  • Humidity: Functions within specification after 24 hour exposure to 90% humidity and subsequent 6-hour recovery with use of the nitrogen supply.
  • Ambient light exposure: Functions within specifications when exposed to ambient light up to 1,000 Lux with cover on the reader.
  • Shock: Withstands a 1-cm drop onto a concrete surface.

Ordering Alerts:

Computer requirements: Any commercially available PC system with Windows™ 7, XP, 2000, or NT 4. 32 bit OS only. RS232 strongly recommended. (Please contact your sales representative for approved USB to RS232 converters).

Compatible with:

Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories

Recommended for:

Treatment planning verification, diagnostic dose studies, quality assurance, whole body dosimetry, extremity dosimetry, environmental dosimetry, high dose applications (radiation hardness testing), and many more.

Figuras

Documentos y descargas

Certificados

    Preguntas frecuentes

    Citas y referencias

    Search citations by name, author, journal title or abstract text