Especificaciones
Consumo de energía4 amps
DescripciónPXS5-928 Microfocus X-Ray Source
Diámetro de la ventana (imperial)Approximately 78 in.
Diámetro de la ventana (métrico)Approximately 20 mm
Rango de humedadAmbient Temperature and Humidity 0 to 32°C, 0-95% RH, up to 5,000 feet
Corriente máx. del haz.160 mA @ 50kV
Consumo máx.8 watts, 50-90kV
Intervalo de tensiones de funcionamiento20-90kV
Tipo de blindajeX-Ray leakage behind the X-Ray tube is less than 0.5mR/hour, measured one inch away with Victoreen 190
Elipticidad de mancha±10% @ 8 watts, 90kV (either axis referred to average)
Espaciado de la mancha a la ventana6 mm (.236 in.)
Material objetivoTungsten
Peso (imperial)Approximately 8 lb.
Peso (métrico)Approximately 3.6 kg
Detalles del módulo de controlDigital control and monitoring of operating conditions and status through RS-232-C port
Potencia de entrada12-14 VDC, 4A
Método de refrigeraciónInternal fan is sufficient for ambient temperature up to 28°C. Adequate air circulation around unit must be
provided. Above 28°C ambient, cooling air must be directed at the unit.
Estabilidad de la ubicación de la manchaTypically <10 microns in 4 hours after 10-minute warm up
Tamaño de mancha4 μ @ 2 watts, 50 to 90kV
5 μ @ 4 watts, 50 to 90kV
9 μ @ 8 watts, 50 to 90kV
Window MaterialBeryllium
Cone of Illumination108° Nominal
Unit SizeEach