Thermo Scientific™

Continuous Ambient Particulate TEOM™ Monitor, Series 1400ab

Número de catálogo: 1400AB
Thermo Scientific™

Continuous Ambient Particulate TEOM™ Monitor, Series 1400ab

Número de catálogo: 1400AB

This product has been discontinued however we continue to provide services and support for this product.

 
Número de catálogo
1400AB
Formato
Each
Descripción
Model TEOM 1400ab Amibent Particulate Monitor
Precio (CLP)
Especificaciones completas
Caudal0.5 to 4.0 L/min. (Main), 2.0 to 18.0 L/min. (Auxiliary)
Entradas7 Averaged Analog Inputs (±2 VDC or ±10 VDC) with User-defined Conversion to Engineering Units, including vector-averaged wind velocity and direction
Método de mediciónTapered Element Oscillating Microbalance (TEOM) Technology
Temperatura de funcionamiento (imperial)35.6°F to 104°F
Temperatura de funcionamiento (métrico)2°C to 40°C
Exactitud±0.75%
Amperaje1/0.5 A (Sensor and Control Units); 4.25/2.25 A (Pump)
Certificaciones/conformidadUSEPA equivalency designation number EQPM-1090-079 as an equivalent PM-10 monitor, USEPA CACM for PM-2.5German EPA approval for TSP, PM-10; conforms with European continuous PM-10 monitoring requirementsNorms/approvals in Australia, Japan, Korea, and Taiwan; CE EN550011 Group 1, Class B (Emissions), EN55082-1 (Immunity), EN61010-1 (Safety); ETL UL- and CSA-equivalent approval
Puntos de datosInternal datalogging of 1 to 8 user-specified variables; capacity of 40 weeks of hourly mass concentration data
Fondo (imperial)11 in.
Fondo (métrico)28 cm
DescripciónModel TEOM 1400ab Amibent Particulate Monitor
Frecuencia50/60 Hz
Altura (imperial)9 in.
Altura (métrico)22 cm
Intervalo de medición0 to 5,000,000 μg/cu. m (5 g/cu. m)
Salidas3 user-defined analog outputs (0 to 1, 0 to 2, or 0 to 10VDC); 2 User-defined contact closure alarm circuits; two-way RS-232 communication using AK Protocol or German Ambient Network Protocol; and four-line display on TEOM Control Unit
Precisión±1.5 μg/cu. m (1-hr.), ± 0.5μg/cu. m (24-hr.)
Resolución0.1μg/m3
Tensión120/240 V
Peso (imperial)40 lb. (Base of TEOM Sensor Unit)/32 lb. (TEOM Control Unit)
Peso (métrico)18 kg (Base of TEOM Sensor Unit)/15 kg (TEOM Control Unit)
Anchura (imperial)17 in.
Anchura (métrico)43 cm
Unit SizeEach
Mostrando 1 de 1
Número de catálogoEspecificacionesFormatoDescripciónPrecio (CLP)
1400ABEspecificaciones completas
EachModel TEOM 1400ab Amibent Particulate MonitorContáctenos ›
Caudal0.5 to 4.0 L/min. (Main), 2.0 to 18.0 L/min. (Auxiliary)
Entradas7 Averaged Analog Inputs (±2 VDC or ±10 VDC) with User-defined Conversion to Engineering Units, including vector-averaged wind velocity and direction
Método de mediciónTapered Element Oscillating Microbalance (TEOM) Technology
Temperatura de funcionamiento (imperial)35.6°F to 104°F
Temperatura de funcionamiento (métrico)2°C to 40°C
Exactitud±0.75%
Amperaje1/0.5 A (Sensor and Control Units); 4.25/2.25 A (Pump)
Certificaciones/conformidadUSEPA equivalency designation number EQPM-1090-079 as an equivalent PM-10 monitor, USEPA CACM for PM-2.5German EPA approval for TSP, PM-10; conforms with European continuous PM-10 monitoring requirementsNorms/approvals in Australia, Japan, Korea, and Taiwan; CE EN550011 Group 1, Class B (Emissions), EN55082-1 (Immunity), EN61010-1 (Safety); ETL UL- and CSA-equivalent approval
Puntos de datosInternal datalogging of 1 to 8 user-specified variables; capacity of 40 weeks of hourly mass concentration data
Fondo (imperial)11 in.
Fondo (métrico)28 cm
DescripciónModel TEOM 1400ab Amibent Particulate Monitor
Frecuencia50/60 Hz
Altura (imperial)9 in.
Altura (métrico)22 cm
Intervalo de medición0 to 5,000,000 μg/cu. m (5 g/cu. m)
Salidas3 user-defined analog outputs (0 to 1, 0 to 2, or 0 to 10VDC); 2 User-defined contact closure alarm circuits; two-way RS-232 communication using AK Protocol or German Ambient Network Protocol; and four-line display on TEOM Control Unit
Precisión±1.5 μg/cu. m (1-hr.), ± 0.5μg/cu. m (24-hr.)
Resolución0.1μg/m3
Tensión120/240 V
Peso (imperial)40 lb. (Base of TEOM Sensor Unit)/32 lb. (TEOM Control Unit)
Peso (métrico)18 kg (Base of TEOM Sensor Unit)/15 kg (TEOM Control Unit)
Anchura (imperial)17 in.
Anchura (métrico)43 cm
Unit SizeEach
Mostrando 1 de 1
The TEOM Monitor, Series 1400ab, measures PM-10 or PM-2.5 mass concentrations and consists of a TEOM mass sensor and control unit in a network ready configuration. The TEOM 1400ab distinguishes itself from other PM measurement methods by utilizing a direct mass measurement that is not subject to measurement uncertainties found in other surrogate technologies. The TEOM 1400ab provides a self-referencing, NIST traceable direct mass measurement.

Key Features:

  • Utilizes TEOM technology to provide direct mass measurements
  • U.S. and internationally approved particulate monitor
  • Configurable to measure PM-10, PM-2.5, PM-1 or TSP concentrations
  • Easily accommodates all sighting requirements
  • Controller provides for internal data storage with digital and analog input/output capabilities

Standard System Configuration:

  • TEOM sensor and control units
  • Menu-driven software for user interaction via the keypad
  • ActiVol flow control using Automatic Mass Flow Controller(s) and Ambient Temperature Pressure Sensors
  • RPCOMM software for local or remote communication
  • Support for 8500 FDMS and/or ACCU™ System

Includes:

  • Connecting and interface cables and vacuum pump Pallflex TX40 filter, 13mm effective diameter

Compatible with:

  • 8500 FDMS
  • ACCU
  • Streamline Pro MultiCal System
  • Complete Outdoor Enclosure
  • Sample Equilibrium System (SES)

Recommended for:

  • Air quality monitoring networks, including background sites
  • Special studies and supersites
  • Routine input for air quality index
  • Remediation projects (Superfund, hazardous waste)
  • Indoor air, exposure chamber, and industrial hygiene measurements
  • In and around industrial and material handling facilities

Figuras

Documentos y descargas

Certificados

    Preguntas frecuentes

    Citas y referencias

    Search citations by name, author, journal title or abstract text