Thermo Scientific™

Partisol™ Plus 2025 Sequential Ambient Particulate Sampler

Catalog number: 2025TS
Thermo Scientific™

Partisol™ Plus 2025 Sequential Ambient Particulate Sampler

Catalog number: 2025TS

This product has been discontinued however we continue to provide services and support for this product.

 
Catalog Number
2025TS
also known as 2025
Unit Size
Each
Price (EUR)
Full specifications
Flow Rate5 to 18 L/min.
Temperature (English) Operating-22°F to +122°F
Temperature (Metric) Operating-30°C to +50°C
Amperage3/1.5 A
Certifications/CompliancePM-2.5: USEPA Reference Method RFPS-0498-118. PM-10: USEPA Reference Method RFPS-1298-127. EN12341: Equivalent PM-10 manual sampling method for European Norm EN1241. Meets: CE EN550011 Group 1, Class B (Emissions), EN55082-1 (immunity), EN61010-1 (Safety)
Data Points16 days of interval data (stored every 5 minutes), 32 days of input data (stored every 30 minutes, default), and 50 filter records
DescriptionPartisol-Plus 2025 Sequential Air Sampler
OutputsOutputs: Keypad/display for data retrieval and user programmingRS-232 interface for data retrieval and remote operation; supports advanced bi-drectional AK Protocol3 user-defined analog outputs (0 to 5VDC) for datalogging2 user-defined TTL alarm outputsWind vane/anemometer connection with 24 Vdc power output.
Inputs: 2 0-5 Vdc inputs for wind speed and direction. 3 averaged analog inputs (0-5 Vdc) with conversion to engineering units (in addition to wind speed and direction above). RS485 interface for connection with other Thermo Scientific devices.
Voltage120/240 V
Weight (English)101 lb.
Weight (Metric)46 kg
Unit SizeEach
Showing 1 of 1
Catalog NumberSpecificationsUnit SizePrice (EUR)
2025TS
also known as 2025
Full specifications
EachContact Us ›
Flow Rate5 to 18 L/min.
Temperature (English) Operating-22°F to +122°F
Temperature (Metric) Operating-30°C to +50°C
Amperage3/1.5 A
Certifications/CompliancePM-2.5: USEPA Reference Method RFPS-0498-118. PM-10: USEPA Reference Method RFPS-1298-127. EN12341: Equivalent PM-10 manual sampling method for European Norm EN1241. Meets: CE EN550011 Group 1, Class B (Emissions), EN55082-1 (immunity), EN61010-1 (Safety)
Data Points16 days of interval data (stored every 5 minutes), 32 days of input data (stored every 30 minutes, default), and 50 filter records
DescriptionPartisol-Plus 2025 Sequential Air Sampler
OutputsOutputs: Keypad/display for data retrieval and user programmingRS-232 interface for data retrieval and remote operation; supports advanced bi-drectional AK Protocol3 user-defined analog outputs (0 to 5VDC) for datalogging2 user-defined TTL alarm outputsWind vane/anemometer connection with 24 Vdc power output.
Inputs: 2 0-5 Vdc inputs for wind speed and direction. 3 averaged analog inputs (0-5 Vdc) with conversion to engineering units (in addition to wind speed and direction above). RS485 interface for connection with other Thermo Scientific devices.
Voltage120/240 V
Weight (English)101 lb.
Weight (Metric)46 kg
Unit SizeEach
Showing 1 of 1
The Partisol Plus 2025 measures PM-10, PM-2.5, or PM-1 on 47mm filter cassettes contained in removable magazine cartridges. The 2025 is a performance-based stand alone unit with a proven track record in the field that operates in a wide variety of climates and meets the US EPA guidelines for manual air samplers.

Key Features:

  • U.S. EPA PM-2.5 and PM-10 reference method designation
  • Sequential sampling with a 16-filter capacity provides two weeks of unattended daily sampling
  • Device exchanges 47mm diameter sample filter automatically by a user-defined time interval or other condition
  • Easily configurable for PM-10, PM-2.5 or PM-1 using a Very Sharp Cut Cyclone (VSCC), Sharp Cut Cyclone (SCC), WINS Impactor or a downtube
  • Sampler is designed to be installed at an outdoor sampling location without additional protection

Standard System Configuration:

  • Single unit consisting of manual sampler, stand and sampling inlet
  • ActiVol automatic flow control
  • RPCOMM software for local and remote interaction
  • Storage magazines with 16 filter capacity
  • Standard 47mm filter cassettes with convenient filter change mechanism

Includes:

  • Consumables for average first year's operation
  • Sampler, Stand, Pump, and Accessories

Compatible with:

  • Streamline Pro
  • Streamline FTS

Recommended for:

  • Air quality monitoring networks, including background sites
  • Special studies and super-sites
  • Routine input for air quality index
  • In and around industrial and material handling facilities
  • Remediation projects (Superfund, hazardous waste)

Figures