Specifications
Format96-well Plate
Thermal Accuracy0.25°C (35°C to 95°C after 3 min.)
Sample Ramp Rate± 1.1°C⁄s
Sensitivity1 Copy of RNase P gene (Human Genomic DNA)
Temperature Range (Metric)4 to 100°C
Reaction Volume Range20–100 μL
Passive Reference DyeROX (Separate Tube), No ROX, ROX (Pre-mixed)
Dynamic Range9 Logs
Calibrated DyeSYBR™ Green I, ROX™, NED™, TAMRA™, JOE™, Texas Red™, Cy5™, VIC™, FAM™, Cy3™
High-throughput CompatibilityMultiplexing
For Use With (Equipment)7500 System
Thermal Uniformity0.50°C (after 30 s)
Reaction SpeedStandard
Throughput4 to 5-96-well Plates per 8-hr day
OpticsCCD Camera, 5 Excitation Filters, Tungsten-halogen Lamp, 5 Emission Filters
Thermal Cycling SystemPeltier-based System
DimensionsWidth 34 cm (13.39 in.), Depth 45 cm (17.72 in.), Height 49 cm (19.29 in.)
External Computer80 GB HD, Dell Minitower: Celeron 346 (3.06 GHz), 1 GB NON ECC 533 MHz DDR2, 16X DVD
Software CompatibilityPrimer Express, 21 CFR Part 11 Module (optional), 7500 System Software (RQ included)
Precision99.7% Confidence (5000 & 10000 Copies)
Voltage120 V
Electrical Requirements950 W
Peak Block Ramp Rate2.5°C⁄s
Max. Heating Rate (Thermal Block)2.5°C
Quantity1 system
Weight34 kg (75 lb.)
Unit SizeEach