Thermo Scientific™

FHZ 691-10 Dose Rate Detector

Catalog number: 4254271
Thermo Scientific™

FHZ 691-10 Dose Rate Detector

Catalog number: 4254271
Detect even the smallest amounts of artificial radioactivity with a high sensitivity scintillation detector. The Thermo Scientific™ FHZ 691-10 probe is specially suited to detect dose rate increases as small as 3 percent at normal background within a measuring time of one minute. Using the Natural Background Rejection (NBR) algorithm, this low dose rate detector easily differentiates between artificial parts of radiation and fluctuations in the natural dose rate (e.g. caused by precipitation) in the surrounding environment.
 
Catalog Number
4254271
Unit Size
Each
Description
FHZ 691-10 Dose Rate Detector
Price (HKD)
Full specifications
DescriptionFHZ 691-10 Dose Rate Detector
Detector TypeScintillation
Measurement RangeH*(10): Low dose detector NaI(Tl): 0.01 μSv/h to 0,5 mSv/h
High dose detector (GM): 0.5 mSv/h to 10 Sv/h
SensitivityLow dose detector: 160 cps/(μSv/h)
High dose detector : 17 cps/(mSv/h)
Weight (Metric)520 g
Energy RangeH*(10): 60 keV to 3 MeV
RangeAngular: - 45° bis 45°
Storage Temperature (Metric)- 40°C bis + 70°C
Product TypeDose Rate Detector
Dimension (Metric)Diameter: 65 mm
Length: 257 mm (without plug-in connection)
ClassProtection class: IP67 in accordance DIN 40050/EN60529
Permissible Detector Dose10 Sv
Operating Limit Temperature Range (Metric)- 30°C bis + 55°C
Unit SizeEach
Showing 1 of 1
Catalog NumberSpecificationsUnit SizeDescriptionPrice (HKD)
4254271Full specifications
EachFHZ 691-10 Dose Rate DetectorRequest A Quote
DescriptionFHZ 691-10 Dose Rate Detector
Detector TypeScintillation
Measurement RangeH*(10): Low dose detector NaI(Tl): 0.01 μSv/h to 0,5 mSv/h
High dose detector (GM): 0.5 mSv/h to 10 Sv/h
SensitivityLow dose detector: 160 cps/(μSv/h)
High dose detector : 17 cps/(mSv/h)
Weight (Metric)520 g
Energy RangeH*(10): 60 keV to 3 MeV
RangeAngular: - 45° bis 45°
Storage Temperature (Metric)- 40°C bis + 70°C
Product TypeDose Rate Detector
Dimension (Metric)Diameter: 65 mm
Length: 257 mm (without plug-in connection)
ClassProtection class: IP67 in accordance DIN 40050/EN60529
Permissible Detector Dose10 Sv
Operating Limit Temperature Range (Metric)- 30°C bis + 55°C
Unit SizeEach
Showing 1 of 1

The FHZ 691-10 intelligent probe is used for H*(10) ambient equivalent dose rate measurements over a dynamic range of 9 decades. 

  • High measuring accuracy, 3% at 100 nSv/h in 1 min.
  • ADF filter for filtering statistical fluctuations and fast reaction to changes. 
  • PTB type approval for calibration in combination with the display unit FHT 6020P. 
  • RS-485 interface. 
  • Transfer protocol with very short reaction times for monitoring networks. 
  • Extended self checks for probe functions.
  • Surge protection for all inputs and outputs.

Figures

Documents & Downloads

Certificates

    Frequently asked questions (FAQs)

    Citations & References

    Search citations by name, author, journal title or abstract text