規格
DescriptionPSM-500 On-line Particle Size Analyzer
MaterialWetted parts: engineering polymers or rubber covered steel
Surface CoatingCommercial sandblast, surface preparation, polyurethane fill primer two part epoxy color finish
EnvironmentDirect exposure to sunlight and ultraviolet radiation is not permitted. Not rated for unprotected, out-of-doors operation
Instrument AirMinimum Pressure - 552 kPa; Maximum Pressure - 1034 kPa; Maximum Flow - 17m3/h @ 690kPa
Percent Solids% Solids Wt 10% Minimum to 60% Maximum, % Solids Vol 3% Minimum to 26% Maximum
Quality AssuranceAdelaide manufacturing facility ISO-9001:2000 certified
Sample TypeChemical: Caustic slurries up to PH of 12.5
Standardization WaterClean potable, Flow 6 to 18 L/Day
VibrationNot to exceed 10G at 20Hz
Water for Sample AcquisitionMinimum Pressure 276 kPa, Maximum Pressure 689 kPa, Flow 45 L/min.
Density2.0 to 5.5 SG for dry solids
DiameterMaximum Transport Diameter: <2.3 mm @ <1% retained by Wt. @ 2.7 SG, <1.2 mm @ <1% retained by Wt. @ 5.1 SG
Operating TemperatureAmbient -10°C to +50°C Operating Altitude: 0 to 5000 mts
Electrical Requirements380/600 V 50/60 Hz 3 Phase
Electronics Connection to AnalyzerCAT5/6/6e network cable directly to analyzer (less than 100m in length)
Electronics EnclosureIP65; CE; ICES 001, CSA, FCC Part 15B and RCM
Unit SizeEach