Search Thermo Fisher Scientific
Search Thermo Fisher Scientific
Please enter the following values to determine the percentage of PCR products containing an error when amplified with Thermo Scientific Phusion High-Fidelity DNA Polymerase |
Length of PCR product in bp |
Fidelity value of DNA polymerase Phusion High-Fidelity DNA polymerase (HF Buffer; fidelity 4.4 x 10-7)Phusion High-Fidelity DNA polymerase (GC Buffer; fidelity 9.5 x 10-7) Pyrococcus furiosus DNA polymerase (2.8 x 10-6) Taq DNA polymerase (2.28 x 10-5) |
Number of PCR cycles |
Estimated percentage of PCR products having an error (i.e., DNA molecules with 1 error): |
Length of PCR product in bp: |
Number of PCR cycles: |
What percent of the product molecules contain an error after PCR (30 cycles) with different polymerases?
Polymerase | 1 kb template | 3 kb template |
---|---|---|
Phusion High-Fidelity DNA Polymerases (HF Buffer) | 1.32% | 3.96% |
Phusion High-Fidelity DNA Polymerases (GC Buffer) | 2.85% | 8.55% |
Pyrococcus furiosus DNA polymerase | 8.4% | 25.2% |
Taq DNA polymerase | 68.4% | 205.2% |
The table above demonstrates the low error rate of Phusion DNA Polymerase. After 30 cycles of PCR amplifying a 3 kb template, only 3.96 % of the product DNA molecules contain 1 (nucleotide) error each. This means that 96.04 % of the product molecules are entirely error-free. In contrast, after the same PCR protocol performed with Taq DNA polymerase, every product molecule contains an average of 2 errors.