Specifications
Amperage15 A
Capillary Length50 cm, 36 cm
Electrical Requirements100 V to 240 V, 50/60 HZ, 15 A
External ComputerHard Drive: 2x 500GB SATA 3.0 Gb/s and 8 MB Data Burst Cache, Memory: 16 GB (2x 8GB) 1600 MHz DDR3 Non-ECC. 4th Gen Intel Core I7 Processor- 3.1 GHz Turbo Processor, Operating System: Windows 7 SP1
For Use With (Equipment)3500 Genetic Analyzer Capillary Array (50 cm), POP-6™ Polymer for 3500⁄3500xL Genetic Analyzers, POP-4™ Polymer for 3500⁄3500xL Genetic Analyzers, 3500 Genetic Analyzer Capillary Array (36 cm), POP-7™ Polymer for 3500⁄3500xL Genetic Analyzers
Format96-well Plate, 8-tube Strip
Height (English) Exterior28.34 in.
Height (Metric) Exterior72 cm
Length (English) Exterior24.01 in.
Length (Metric) Exterior61 cm
Max. VoltageElectrophoresis Voltage - Up to 20 kV
Number of Capillaries8 Capillaries
Operating EnvironmentHumidity: 20%-80% (non-condensing), Room temp: <2° fluctuation during runs, Temperature: 15°C-30°C
Operating SystemWindows 7 (Service Pack 1)
PolymerPOP-4™, POP-6™, POP-7™
Polymer ConsumptionReplace at earliest of 7 days⁄384 samples⁄60 Injections (384-sample pouches), Replace at earliest of 7 days⁄960 samples⁄120 Injections (960-sample pouches)
Quantity1 System
Read Length≥850 bp using a 50 cm capillary and POP-7™
Software CompatibilitySequencing Analysis Software, Variant Reporter™ Software, GeneMapper™ Software
ThroughputUp to 424 samples per day (using run module HID36 POP-7)
Warranty1 Year Limited Warranty on Parts and Labor, Service Installation and Application Training Included
Weight82 kg (approximately)
Width (English) Exterior24.01 in.
Width (Metric) Exterior61 cm
Software TypeData Collection Software, Fragment Analysis
Starting MaterialDNA