Thermo Scientific™

Vanquish™ Fluorescence Detectors

Catalog number: VF-D50-A
Thermo Scientific™

Vanquish™ Fluorescence Detectors

Catalog number: VF-D50-A

Get trace detection performance through highly effective stray light suppression with Thermo Scientific™ Vanquish™ Fluorescence Detectors. These detectors reduce thermal effects using temperature-controlled flow cells for increased detection precision. Monitor up to four wavelength pairs simultaneously, or scan your chromatogram for the best emission/excitation wavelengths. Improve detection sensitivity and selectivity by leveraging ultrafast wavelength switching between peaks.

View Fluorescence Detector Resources

 
Catalog Number
VF-D50-A
Unit Size
Each
Description
Vanquish™ Fluorescence Detector F, single PMT
Wavelength Range
Ex Min 200; Ex Max 630; Em Min 220; Em Max 650
Price (BRL)
Full specifications
DescriptionVanquish™ Fluorescence Detector F, single PMT
Data Collection Rate200 Hz Max.
Detection ModeUp to 4 Channels
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
FilterEmission Filter: Variable, 5 positions
Flow CellStandard flow cell: 8 μL, biocompatible; 20 bar (290 psi) maximum cell pressure
Micro flow cell: 2 μL, biocompatible; 40 bar (290 psi) maximum cell pressure
Flow Cell ThermostattingAmbient +15°C to 50°C
For Use With (Equipment)All Vanquish UHPLC Systems
GLP FeaturesPredictive Performance functions: lamp age, leak detection, service monitoring period, grating and filter movements, PMT workload. All system parameters logged in the Chromeleon Chromatography Data System (CDS) audit trail.
LampXenon Flash Lamp
Lamp Pulse FrequencyHighPower (∼300 Hz); Standard (∼100 Hz); LongLife (∼20 Hz)
MaterialBiocompatible (with biocompatible flow cells); pH range 2-12, chloride concentration up to 1 mol/L
No. of Fluorescence ChannelsUp to 4
Optical SystemTwo monochromators w/ concave holographic gratings and elliptic mirrors for highest efficiency in light transmission
PC ConnectionUSB 2.0; 3-port HUB to connect further Vanquish modules
Safety FeaturesPower-up diagnostics of optics, cooling fans, motors and electronics. Leak detection and safe leak handling.
SensitivityRaman S/N: >550 ASTM over the entire lifetime of the lamp (>2100 using dark signal as noise reference)
Spectral BandwidthFWHM: Excitation: 20 nm; Emission: 20 nm
Temperature Range5°C to 35°C; 20 to 80% RH (non condensing)
Wavelength AccuracyWavelength Accuracy: ±2 nm
Wavelength Repeatability; ±0.2 nm
Wavelength RangeEx Min 200; Ex Max 630; Em Min 220; Em Max 650
Dimension(h x w x d) 159 mm x 420 mm x 620 mm (6.3 in. x 16.5 in. x 24.4 in.)
Weight (Metric)21 kg
Electrical Requirements100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Unit SizeEach
Catalog Number
Unit Size
Each
Description
Vanquish™ Fluorescence Detector C
Wavelength Range
Excitation: 200-630 nm, Emission: 220-650 nm
Price (BRL)
Full specifications
DescriptionVanquish™ Fluorescence Detector C
Data Collection Rate100 Hz Max.
Detection ModeSingle Channel
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
Optical SystemTwo monochromators with concave holographic gratings and elliptic mirrors for highest efficiency in light transmission
PC ConnectionUSB 2.0; 3-port hub to connect further Vanquish modules
Wavelength RangeExcitation: 200-630 nm, Emission: 220-650 nm
Dimension159 mm × 420 mm × 620 mm (6.3 in × 16.5 in × 24.4 in)
Weight (Metric)21 kg
Voltage100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Unit SizeEach
Catalog Number
Unit Size
Each
Description
Vanquish™ Fluorescence Detector F, with dual-PMT
Wavelength Range
Dual-PMT
Ex Min 200; Ex Max 880; Em Min 220; Em Max 900
Price (BRL)
Full specifications
DescriptionVanquish™ Fluorescence Detector F, with dual-PMT
Data Collection Rate200 Hz Max.
Detection ModeUp to 4 Channels
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
FilterEmission Filter: Variable, 5 positions
Flow CellStandard flow cell: 8 μL, biocompatible; 20 bar (290 psi) maximum cell pressure
Micro flow cell: 2 μL, biocompatible; 40 bar (290 psi) maximum cell pressure
Flow Cell ThermostattingAmbient +15°C to 50°C
For Use With (Equipment)All Vanquish UHPLC Systems
GLP FeaturesPredictive Performance functions: lamp age, leak detection, service monitoring period, grating and filter movements, PMT workload. All system parameters logged in the Chromeleon Chromatography Data System (CDS) audit trail.
LampXenon Flash Lamp
Lamp Pulse FrequencyHighPower (∼300 Hz); Standard (∼100 Hz); LongLife (∼20 Hz)
MaterialBiocompatible (with biocompatible flow cells); pH range 2-12, chloride concentration up to 1 mol/L
No. of Fluorescence ChannelsUp to 4
Optical SystemTwo monochromators w/ concave holographic gratings and elliptic mirrors for highest efficiency in light transmission
PC ConnectionUSB 2.0; 3-port HUB to connect further Vanquish modules
Safety FeaturesPower-up diagnostics of optics, cooling fans, motors and electronics. Leak detection and safe leak handling.
SensitivityRaman S/N: >550 ASTM over the entire lifetime of the lamp (>2100 using dark signal as noise reference)
Spectral BandwidthFWHM: Excitation: 20 nm; Emission: 20 nm
Temperature Range5°C to 35°C; 20 to 80% RH (non condensing)
Wavelength AccuracyWavelength Accuracy: ±2 nm
Wavelength Repeatability; ±0.2 nm
Wavelength RangeDual-PMT
Ex Min 200; Ex Max 880; Em Min 220; Em Max 900
Dimension(h x w x d) 159 mm x 420 mm x 620 mm (6.3 in. x 16.5 in. x 24.4 in.)
Weight (Metric)21 kg
Electrical Requirements100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Unit SizeEach
Catalog Number
Unit Size
Each
Description
Vanquish™ Fluorescence Detector C with Dual-PMT
Wavelength Range
Excitation: 200-880 nm, Emission: 220-900 nm
Price (BRL)
Full specifications
DescriptionVanquish™ Fluorescence Detector C with Dual-PMT
Data Collection Rate100 Hz Max.
Detection ModeSingle Channel
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
Optical SystemTwo monochromators with concave holographic gratings and elliptic mirrors for highest efficiency in light transmission
PC ConnectionUSB 2.0; 3-port hub to connect further Vanquish modules
Wavelength RangeExcitation: 200-880 nm, Emission: 220-900 nm
Dimension159 mm × 420 mm × 620 mm (6.3 in × 16.5 in × 24.4 in)
Weight (Metric)21 kg
Voltage100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Unit SizeEach
Showing 4 of 4
Catalog NumberSpecificationsUnit SizeDescriptionWavelength RangePrice (BRL)
VF-D50-AFull specifications
EachVanquish™ Fluorescence Detector F, single PMTEx Min 200; Ex Max 630; Em Min 220; Em Max 650Contact Us ›
DescriptionVanquish™ Fluorescence Detector F, single PMT
Data Collection Rate200 Hz Max.
Detection ModeUp to 4 Channels
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
FilterEmission Filter: Variable, 5 positions
Flow CellStandard flow cell: 8 μL, biocompatible; 20 bar (290 psi) maximum cell pressure
Micro flow cell: 2 μL, biocompatible; 40 bar (290 psi) maximum cell pressure
Flow Cell ThermostattingAmbient +15°C to 50°C
For Use With (Equipment)All Vanquish UHPLC Systems
GLP FeaturesPredictive Performance functions: lamp age, leak detection, service monitoring period, grating and filter movements, PMT workload. All system parameters logged in the Chromeleon Chromatography Data System (CDS) audit trail.
LampXenon Flash Lamp
Lamp Pulse FrequencyHighPower (∼300 Hz); Standard (∼100 Hz); LongLife (∼20 Hz)
MaterialBiocompatible (with biocompatible flow cells); pH range 2-12, chloride concentration up to 1 mol/L
No. of Fluorescence ChannelsUp to 4
Optical SystemTwo monochromators w/ concave holographic gratings and elliptic mirrors for highest efficiency in light transmission
PC ConnectionUSB 2.0; 3-port HUB to connect further Vanquish modules
Safety FeaturesPower-up diagnostics of optics, cooling fans, motors and electronics. Leak detection and safe leak handling.
SensitivityRaman S/N: >550 ASTM over the entire lifetime of the lamp (>2100 using dark signal as noise reference)
Spectral BandwidthFWHM: Excitation: 20 nm; Emission: 20 nm
Temperature Range5°C to 35°C; 20 to 80% RH (non condensing)
Wavelength AccuracyWavelength Accuracy: ±2 nm
Wavelength Repeatability; ±0.2 nm
Wavelength RangeEx Min 200; Ex Max 630; Em Min 220; Em Max 650
Dimension(h x w x d) 159 mm x 420 mm x 620 mm (6.3 in. x 16.5 in. x 24.4 in.)
Weight (Metric)21 kg
Electrical Requirements100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Unit SizeEach
VC-D50-A-01Full specifications
EachVanquish™ Fluorescence Detector CExcitation: 200-630 nm, Emission: 220-650 nmRequest A Quote
DescriptionVanquish™ Fluorescence Detector C
Data Collection Rate100 Hz Max.
Detection ModeSingle Channel
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
Optical SystemTwo monochromators with concave holographic gratings and elliptic mirrors for highest efficiency in light transmission
PC ConnectionUSB 2.0; 3-port hub to connect further Vanquish modules
Wavelength RangeExcitation: 200-630 nm, Emission: 220-650 nm
Dimension159 mm × 420 mm × 620 mm (6.3 in × 16.5 in × 24.4 in)
Weight (Metric)21 kg
Voltage100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Unit SizeEach
VF-D51-AFull specifications
EachVanquish™ Fluorescence Detector F, with dual-PMTDual-PMT
Ex Min 200; Ex Max 880; Em Min 220; Em Max 900
Contact Us ›
DescriptionVanquish™ Fluorescence Detector F, with dual-PMT
Data Collection Rate200 Hz Max.
Detection ModeUp to 4 Channels
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
FilterEmission Filter: Variable, 5 positions
Flow CellStandard flow cell: 8 μL, biocompatible; 20 bar (290 psi) maximum cell pressure
Micro flow cell: 2 μL, biocompatible; 40 bar (290 psi) maximum cell pressure
Flow Cell ThermostattingAmbient +15°C to 50°C
For Use With (Equipment)All Vanquish UHPLC Systems
GLP FeaturesPredictive Performance functions: lamp age, leak detection, service monitoring period, grating and filter movements, PMT workload. All system parameters logged in the Chromeleon Chromatography Data System (CDS) audit trail.
LampXenon Flash Lamp
Lamp Pulse FrequencyHighPower (∼300 Hz); Standard (∼100 Hz); LongLife (∼20 Hz)
MaterialBiocompatible (with biocompatible flow cells); pH range 2-12, chloride concentration up to 1 mol/L
No. of Fluorescence ChannelsUp to 4
Optical SystemTwo monochromators w/ concave holographic gratings and elliptic mirrors for highest efficiency in light transmission
PC ConnectionUSB 2.0; 3-port HUB to connect further Vanquish modules
Safety FeaturesPower-up diagnostics of optics, cooling fans, motors and electronics. Leak detection and safe leak handling.
SensitivityRaman S/N: >550 ASTM over the entire lifetime of the lamp (>2100 using dark signal as noise reference)
Spectral BandwidthFWHM: Excitation: 20 nm; Emission: 20 nm
Temperature Range5°C to 35°C; 20 to 80% RH (non condensing)
Wavelength AccuracyWavelength Accuracy: ±2 nm
Wavelength Repeatability; ±0.2 nm
Wavelength RangeDual-PMT
Ex Min 200; Ex Max 880; Em Min 220; Em Max 900
Dimension(h x w x d) 159 mm x 420 mm x 620 mm (6.3 in. x 16.5 in. x 24.4 in.)
Weight (Metric)21 kg
Electrical Requirements100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Unit SizeEach
VC-D51-A-01Full specifications
EachVanquish™ Fluorescence Detector C with Dual-PMTExcitation: 200-880 nm, Emission: 220-900 nmRequest A Quote
DescriptionVanquish™ Fluorescence Detector C with Dual-PMT
Data Collection Rate100 Hz Max.
Detection ModeSingle Channel
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
Optical SystemTwo monochromators with concave holographic gratings and elliptic mirrors for highest efficiency in light transmission
PC ConnectionUSB 2.0; 3-port hub to connect further Vanquish modules
Wavelength RangeExcitation: 200-880 nm, Emission: 220-900 nm
Dimension159 mm × 420 mm × 620 mm (6.3 in × 16.5 in × 24.4 in)
Weight (Metric)21 kg
Voltage100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Unit SizeEach
Showing 4 of 4
  • Acquire data at up to 200 Hz for best support of even the fastest UHPLC separations
  • Tools to optimize analysis sensitivity
  • Simplify method development or improve existing methods using single spectrum scans or fluorescence field acquisition in excitation, emission, or synchronous mode
  • Improve your cost of ownership with increased lamp lifetime offered by our long-life xenon flash lamp and various lamp operation modes
  • Dual-PMT options to extend wavelength range

Figures

Documents & Downloads

Certificates

    Frequently asked questions (FAQs)

    Citations & References

    Search citations by name, author, journal title or abstract text