製品番号(カタログ番号) | 数量 |
---|---|
Q33222 | 275 assays |
Q33221 | 75 assays |
New Qubit 4 and Qubit Flex firmware features include importing sample IDs, renaming files before exporting, a countdown timer, and on-board troubleshooting guidance. Get the most out of your Qubit assays with these powerful new tools.
Qubit RNA IQ Assay Kits are not compatible with plate readers. The IQ value calculations are performed by the Qubit instrument, which involves proprietary algorithms that cannot be replicated on other devices. For accurate results, use the Qubit instrument as recommended.
Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.
There is no simple way to convert the RNA IQ score to RIN, because they measure different things. The application note, Qubit RNA IQ Assay: a fast and easy fluorometric RNA quality assessment, shows a detailed comparison and explanation of the two measurements. Figures 2 and 4 are most relevant.
No. The presence of an excess amount of free nucleotides/bases may affect the assay results in the same manner as other contaminants, but the reagents in this kit do not give a fluorescence response to nucleotides or bases.
Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.
At present, we do not have a fluorescence-based assay that tests for sample purity (protein contamination). We do offer various instruments that can obtain absorbance readings (absorbance at 260/280), such as our NanoDrop, BioMate and GENESYS spectrophotometers.
Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.
No. The Qubit RNA IQ Assay does not assay the purity of the RNA sample, only the extent of degradation of the RNA (small versus large strands).
Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.